Title :
A High Rate Proportional Chamber
Author :
Henderson, R. ; Faszer, W. ; Openshaw, R. ; Sheffer, G. ; Salomon, M. ; Dew, S. ; Marans, J. ; Wilson, P.
Author_Institution :
TRIUMF, 4004 Wesbrook Mall, Vancouver, B. C., Canada, V6T-2A3
Abstract :
Gas mixtures with high specific ionization allow the use of small interelectrode distances while still maintaining full efficiency. With the short electron drift distances the timing resolution is also improved. We have built and operated two 25 cm2 chambers with small interelectrode distances. Also single wire detector cells have been built to test gas mixture lifetimes. Various admixtures of CF4, DME, Isobutane, Ethane and Argon have been tested. Possible applications of such chambers are as beam profile monitors, position tagging of rare events and front end chambers in spectrometers.
Keywords :
Anodes; Capacitance; Cathodes; Detectors; Instruments; Preamplifiers; Spectroscopy; Tagging; Testing; Wire;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1987.4337399