• DocumentCode
    916785
  • Title

    Minimisation of gradient artefacts in cardiac-MRI-gating using adaptive interference cancellation filter with synthesised reference

  • Author

    Park, H.D. ; Cho, S.-P. ; Lee, K.-J. ; Park, Y.C.

  • Author_Institution
    Yonsei Univ., Wonju
  • Volume
    43
  • Issue
    20
  • fYear
    2007
  • Firstpage
    1070
  • Lastpage
    1071
  • Abstract
    A simple and successful method for cardiac-MRI-gating is proposed. The adaptive interference cancellation filter (AICF) is used with a synthesised reference signal to reduce the gradient artefacts caused by the magnetic resonance (MR). The reference signals of the AICF were a combination of the noisy, three-channel ECG signals. In particular, the proposed method is based on a simple experimental setup and does not require any information from amplifiers of the MRI machine, such as shape, amplitude and rise time.
  • Keywords
    biomedical MRI; electrocardiography; image denoising; interference filters; interference suppression; medical image processing; AICF; adaptive interference cancellation filter; cardiac-MRI-gating; gradient artefact minimisation; magnetic resonance imaging; synthesised reference signal; three-channel ECG signal;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:20070760
  • Filename
    4338188