• DocumentCode
    918975
  • Title

    A Validated Model for the Zero Drift Due to Eddy Currents in Electromagnetic Flowmeters Operating With Electrolytic Conductors

  • Author

    Maalouf, Aline I.

  • Author_Institution
    Dargroup, Beirut
  • Volume
    7
  • Issue
    11
  • fYear
    2007
  • Firstpage
    1497
  • Lastpage
    1505
  • Abstract
    In this paper, the author investigates theoretically the magnitudes of the zero offsets and the zero drifts originating from eddy currents in electromagnetic flowmeters operating with electrolytic conductors. The dependence of such zero offsets on the liquid properties, frequency of operation, etc., is explained. This dependence is used to predict the zero offsets expected in metering electrolytic conductors using phase sensitive detection with the flowmeter tube described in this paper. Precautions needed in flowmeter design to minimize instrument offsets and zero drifts are explained. At the end, the proposed model was validated against experimental data.
  • Keywords
    eddy currents; electromagnetic devices; flowmeters; eddy currents; electrolytic conductors; electromagnetic flowmeters; flowmeter tube; impedance spectroscopy; transformer signals; zero drifts; zero offsets; Conductors; Eddy currents; Electrodes; Electromagnetic modeling; Frequency; Geometry; Impedance; Instruments; Magnetic fields; Phase detection; Eddy currents; electromagnetic flowmeter; impedance spectroscopy; transformer signals; zero drift;
  • fLanguage
    English
  • Journal_Title
    Sensors Journal, IEEE
  • Publisher
    ieee
  • ISSN
    1530-437X
  • Type

    jour

  • DOI
    10.1109/JSEN.2007.907562
  • Filename
    4339513