DocumentCode :
922427
Title :
Correction of computed tomography motion artifacts using pixel-specific back-projection
Author :
Ritchie, Cameron J. ; Crawford, Carl R. ; Godwin, J. ; King, Kevin F. ; Kim, Yongmin
Author_Institution :
Dept. of Bioeng., Washington Univ., Seattle, WA, USA
Volume :
15
Issue :
3
fYear :
1996
fDate :
6/1/1996 12:00:00 AM
Firstpage :
333
Lastpage :
342
Abstract :
Cardiac and respiratory motion can cause artifacts in computed tomography scans of the chest. The authors describe a new method for reducing these artifacts called pixel-specific back-projection (PSBP). PSBP reduces artifacts caused by in-plane motion by reconstructing each pixel in a frame of reference that moves with the in-plane motion in the volume being scanned. The motion of the frame of reference is specified by constructing maps that describe the motion of each pixel in the image at the time each projection was measured; these maps are based on measurements of the in-plane motion. PSBP has been tested in computer simulations and with volunteer data. In computer simulations, PSBP removed the structured artifacts caused by motion. In scans of two volunteers, PSBP reduced doubling and streaking in chest scans to a level that made the images clinically useful. PSBP corrections of liver scans were less satisfactory because the motion of the liver is predominantly superior-inferior (S-I). PSBP uses a unique set of motion parameters to describe the motion at each point in the chest as opposed to requiring that the motion be described by a single set of parameters. Therefore, PSBP may be more useful in correcting clinical scans than are other correction techniques previously described
Keywords :
biomechanics; cardiology; computerised tomography; image reconstruction; liver; medical image processing; motion estimation; cardiac motion; clinical scans; computed tomography chest scans; computed tomography motion artifacts correction; computer simulations; correction techniques; doubling; in-plane motion; liver scans; medical diagnostic imaging; motion parameters set; pixel-specific back-projection; respiratory motion; streaking; volunteer data; Biomedical engineering; Biomedical imaging; Computed tomography; Computer simulation; Image reconstruction; Liver; Motion measurement; Pixel; Testing; Time measurement;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.500142
Filename :
500142
Link To Document :
بازگشت