• DocumentCode
    922962
  • Title

    Improved instrument for probing static potential profiles in semiconductor devices

  • Author

    Ladbrooke, P.H.

  • Author_Institution
    Royal Military College of Science, Swindon, UK
  • Volume
    9
  • Issue
    14
  • fYear
    1973
  • Firstpage
    315
  • Lastpage
    316
  • Abstract
    Improvements to the positional accuracy of point-contact probes permit smaller features to be investigated in higher-doped semiconductors. An xy stage with an accuracy of within 0¿5 ¿m has been developed for use with an etched-tungsten probe with a tip radius of about 0.5 ¿m. The probe is constrained to move in a vertical plane by a special cone-pivoted arm so that it can be raised and lowered repeatedly over a fixed point while the sample is moved underheath it. This arm is also counterbalanced so that the probe pressure can be finely controlled. The effectiveness of these measures has been assessed by probing a known (Schottky barrier) potential; an error of 15% over 5 ¿m is typical.
  • Keywords
    point contacts; probes; semiconductor device testing; voltage measurement; GaAs barrier; Ni; etched tungsten probe; point contacts; probes; semiconductor device testing; static potential profiles; voltage measurement;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:19730226
  • Filename
    4236172