• DocumentCode
    923237
  • Title

    Diathermy Applicators with Circular Aperture and Corrugated Flange

  • Author

    Stuchly, Maria A. ; Stuchly, Stanislaw S. ; Kantor, Gideon

  • Volume
    28
  • Issue
    3
  • fYear
    1980
  • fDate
    3/1/1980 12:00:00 AM
  • Firstpage
    267
  • Lastpage
    271
  • Abstract
    A design method and experimental results for a direct-contact circular aperture applicator are provided. The aperture is excited in the TE11 mode; a corrugated flange surrounding the aperture improves the uniformity of the heating pattern and limits leakage. The performance of the applicators operating in the S band (2.45 GHz) and the X band (9.96 GHz) has been tested using a short monopole probe and a thermographic camera. The heating patterns obtained by the two methods are in agreement within the experimental errors. The applicators are suitable for clinical use, as they are lightweight and rugged, and capable of delivering a desired energy dose effectively, thanks to a relatively small standing-wave ratio (SWR < 2) and very low leakage.
  • Keywords
    Apertures; Applicators; Couplings; Curve fitting; Electromagnetic heating; Flanges; Frequency; Microwave devices; Muscles; Skin neoplasms;
  • fLanguage
    English
  • Journal_Title
    Microwave Theory and Techniques, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9480
  • Type

    jour

  • DOI
    10.1109/TMTT.1980.1130054
  • Filename
    1130054