Title :
The Determination of Impedance with a Double-Slug Transformer
Author :
Ellenwood, R.C. ; Hurlburt, E.H.
Author_Institution :
U. S. Department of Commerce, National Bureau of Standards, Washington 25, D.C.
Abstract :
Formulas are derived and techniques given by means of which impedance measurements can be made by the use of a double-slug transformer. The required parameters are the length, spacing, position, and "effective" dielectric constant of the slugs, and the wavelength. A means for the experimental determination of the "effective" dielectric constant is described. The only quantitative measurements required are of length. A qualitative recognition of the electrically matched condition is needed.
Keywords :
Coaxial components; Conductors; Dielectric constant; Dielectric measurements; Frequency; Impedance matching; Impedance measurement; Instruments; Length measurement; Transmission lines;
Journal_Title :
Proceedings of the IRE
DOI :
10.1109/JRPROC.1952.274151