• DocumentCode
    928453
  • Title

    Range Condition and ML-EM Checkerboard Artifacts

  • Author

    You, Jiangsheng ; Wang, Jing ; Liang, Zhengrong

  • Author_Institution
    State Univ. of New York, Stony Brook
  • Volume
    54
  • Issue
    5
  • fYear
    2007
  • Firstpage
    1696
  • Lastpage
    1702
  • Abstract
    The expectation maximization (EM) algorithm for the maximum likelihood (ML) image reconstruction criterion generates severe checkerboard artifacts in the presence of noise. A classical remedy is to impose an a priori constraint for a penalized ML or maximum a posteriori probability solution. The penalty reduces the checkerboard artifacts and also introduces uncertainty because a priori information is usually unknown in clinic. Recent theoretical investigation reveals that the noise can be divided into two components: one is called null-space noise and the other is range-space noise. The null-space noise can be numerically estimated using filtered backprojection (FBP) algorithm. By the FBP algorithm, the null-space noise annihilates in the reconstruction while the range-space noise propagates into the reconstructed image. The aim of this work is to investigate the relation between the null-space noise and the checkerboard artifacts in the ML-EM reconstruction from noisy projection data. Our study suggests that removing the null-space noise from the projection data could improve the signal-to-noise ratio of the projection data and, therefore, reduce the checkerboard artifacts in the ML-EM reconstructed images. This study reveals an in-depth understanding of the different noise propagations in analytical and iterative image reconstructions, which may be useful to single photon emission computed tomography, where the noise has been a major factor for image degradation. The reduction of the ML-EM checkerboard artifacts by removing the null-space noise avoids the uncertainty of using a priori penalty.
  • Keywords
    Radon transforms; computerised tomography; image denoising; image reconstruction; iterative methods; maximum likelihood estimation; medical image processing; probability; attenuated Radon transform; checkerboard artifacts; expectation maximization algorithm; filtered backprojection algorithm; image degradation; image reconstruction criterion; iterative image reconstruction; maximum a posteriori probability solution; maximum likelihood; noise propagation; null-space noise; null-space noise removal; range condition; range-space noise; signal-to-noise ratio; single photon emission computed tomography; Degradation; Image analysis; Image generation; Image reconstruction; Maximum likelihood estimation; Noise generators; Noise reduction; Signal to noise ratio; Single photon emission computed tomography; Uncertainty; Attenuated Radon transform; SPECT; computed tomography; consistent condition; noise analysis;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2007.901198
  • Filename
    4346749