• DocumentCode
    929473
  • Title

    Medical ultrasonic standards at NPL

  • Author

    Preston, R.C. ; Livett, A.J.

  • Author_Institution
    National Physical Laboratory, Division of Radiation Science & Acoustics, Teddington, UK
  • Volume
    131
  • Issue
    4
  • fYear
    1984
  • fDate
    6/1/1984 12:00:00 AM
  • Firstpage
    233
  • Lastpage
    240
  • Abstract
    Absolute measurement of the acoustic output of medical ultrasonic equipment in the megahertz frequency range is discussed, together with the methods developed at NPL for the determination of the total ultrasonic power using a radiation pressure balance and the measurement of spatial and temporal field quantities with miniature piezoelectric hydrophones. The principles of the radiation pressure balance and the absolute calibration of hydrophones, which form the basis of the NPL ultrasonic standards programme, are described. Finally, beam calibration systems, which have been specially developed for the rapid and quantitative evaluation of ultrasonic fields, are briefly discussed.
  • Keywords
    biomedical equipment; biomedical ultrasonics; standards; NPL; acoustic output; beam calibration systems; medical ultrasonic equipment; megahertz frequency range; miniature piezoelectric hydrophones; radiation pressure balance; spatial field quantities; temporal field quantities; total ultrasonic power; ultrasonic standards;
  • fLanguage
    English
  • Journal_Title
    Physical Science, Measurement and Instrumentation, Management and Education - Reviews, IEE Proceedings A
  • Publisher
    iet
  • ISSN
    0143-702X
  • Type

    jour

  • DOI
    10.1049/ip-a-1.1984.0036
  • Filename
    4646180