Title :
Development of a high-frequency (> 50 MHz) copolymer annular-array, ultrasound transducer
Author :
Gottlieb, Emanuel J. ; Cannata, Jonathan M. ; Hu, Chang-Hong ; Shung, K. Kirk
Author_Institution :
Dept. of Biomedial Eng., Southern California Univ., Los Angeles, CA
Volume :
53
Issue :
5
fYear :
2006
fDate :
5/1/2006 12:00:00 AM
Firstpage :
1037
Lastpage :
1045
Abstract :
The development of a high frequency (> 50 MHz) annular array ultrasonic transducer is presented. The array was constructed by bonding a 9 mum P(VDF-TrFE) film to a two-sided polyimide flexible circuit with annuli electrodes on the top layer. Each annulus was separated by a 30 mum kerf and had several electroplated microvias that connected to electrode traces on the bottom side of the flex circuit. In order to improve device sensitivity, each element was electrically matched to an impedance magnitude of 50 Omega and 0deg phase at resonance using a serial inductor and high impedance coaxial cable. The array´s performance was evaluated by measuring the electrical impedance, pulse echo response, and cross talk between elements. The average round trip insertion loss was -33.5 dB after compensating for diffractive and attenuative losses. The measured average center frequency and bandwidth for an element was 55 MHz and 47%, respectively. The measured cross talk between adjacent elements remained below -29 dB at the center frequency in water. A vertical wire phantom was imaged using a single focus transmit beamformer and dynamic focusing receive beamformer. This image showed a significant improvement in lateral resolution over a range of 9 mm after the dynamic focusing receive algorithm was applied. These results correlated well with predictions from a field II simulation. After beamforming, the minimum lateral resolution achieved by the array (-6 dB) was 108 mum at the focus
Keywords :
coaxial cables; crosstalk; electrodes; flexible electronics; image resolution; inductors; polymer blends; ultrasonic transducer arrays; -33.5 dB; 50 ohm; 55 MHz; annuli electrodes; attenuative losses; cross talk; device sensitivity improvement; diffractive loss; dynamic focusing receive beamformer; electroplated microvias; high-frequency copolymer annular-array; impedance coaxial cable; impedance magnitude; minimum lateral resolution; round trip insertion loss; serial inductor; single focus transmit beamformer; two-sided polyimide flexible circuit; ultrasound transducer; vertical wire phantom; Bonding; Electrodes; Focusing; Frequency measurement; Impedance; Polyimides; Pulse measurements; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2006.1632693
Filename :
1632693
Link To Document :
بازگشت