Title :
A New Peak-to-DC Comparator for Audio Frequencies
Author :
Richman, Peter L.
Author_Institution :
Rotek Instrument Corporation, Cambridge, Mass.
Abstract :
An adjustment-free peak-to-dc transfer technique is described for making precision comparisons between audio frequency ac voltages and a reference dc. Limitations and estimated systematic errors are discussed and justifications for an over-all ac-dc transfer worst case error of 50 ppm presented. Photographs of output oscilloscope patterns are shown for various conditions of ac-dc transfer agreement for several different ac sources. Experimental over-all comparison against a modification of the Smith-Clothier peak calibration technique is described with agreement on the order of 30 ppm.
Keywords :
Calibration; Choppers; Frequency; Harmonic distortion; Instruments; Laboratories; Operational amplifiers; Oscilloscopes; Resistors; Voltmeters;
Journal_Title :
Instrumentation, IRE Transactions on
DOI :
10.1109/IRE-I.1962.5006614