• DocumentCode
    937939
  • Title

    N15H3 Double-Beam Maser as a Primary Frequency Standard

  • Author

    De Prins, Jean

  • Author_Institution
    University of Brussels, Brussels, Belgium.
  • Issue
    3
  • fYear
    1962
  • Firstpage
    200
  • Lastpage
    203
  • Abstract
    Ammonia masers are studied concerning their use as frequency standards. They oscillate on the J = K = 3 inversion line of N15H3. Single beam masers allow the realization of a frequency standard with a stability of 2-3.10-11, and an accuracy of about 10-9. Experiments on double beam masers suggest that their use makes it possible to obtain a stability of the order of 10-12 and an accuracy better than 10-10.
  • Keywords
    Hydrogen; Instruments; Magnetic shielding; Masers; Observatories; Stability; Tuning;
  • fLanguage
    English
  • Journal_Title
    Instrumentation, IRE Transactions on
  • Publisher
    ieee
  • ISSN
    0096-2260
  • Type

    jour

  • DOI
    10.1109/IRE-I.1962.5006631
  • Filename
    5006631