• DocumentCode
    938660
  • Title

    A Continuous Comparison at 97 GHz Radiometer

  • Author

    Predmore, C. Read ; Erickson, Neal R. ; Huguenin, G. Richard ; Goldsmith, Paul F.

  • Volume
    33
  • Issue
    1
  • fYear
    1985
  • fDate
    1/1/1985 12:00:00 AM
  • Firstpage
    44
  • Lastpage
    51
  • Abstract
    A continuous comparison radiometer has been implemented at 97 GHz using quasi-optical techniques for local oscillator (LO) injection and realization of a 90° hybrid. Cryogenically cooled Schottky-diode mixers and FET amplifiers give a double-sideband (DSB) system temperature of 250 K. The system is self-calibrating and optimized under computer control. The root-mean-square (rms) fluctuations due to the receiver are less than 0.025 K with a 1-s integration time.
  • Keywords
    Calibration; Fluctuations; Frequency; Local oscillators; Observatories; Radio astronomy; Radiometry; Receivers; Switches; Voltage;