Energy compensated solid state gamma dosimeter
Author
Sinclair, F. ; Clapp, A. ; Entine, G. ; Kronenberg, S.
Author_Institution
RMD, Watertown, MA, USA
Volume
35
Issue
1
fYear
1988
Firstpage
562
Lastpage
566
Abstract
An analytical model of the detector response for a p-i-n diode has been developed and tested with monoenergetic isotopic sources. The model was then used to simulate the detector response over a wide range of photon energies. This matrix was used in turn to calculate a pulse-height correction factor. Irradiation of the device in a variable-energy X-ray beam showed that while there are some fundamental limitations, this approach can improve the accuracy of the measurement of the dose delivered to tissue.<>
Keywords
dosimeters; gamma-ray detection and measurement; semiconductor counters; semiconductor diodes; analytical model; detector response; energy compensated solid state gamma dosimeter; monoenergetic isotopic sources; p-i-n diode; photon energies; pulse-height correction factor; variable-energy X-ray beam; Analytical models; Atomic measurements; Dosimetry; Instruments; Pulse measurements; Radiation detectors; Silicon; Solid state circuits; X-ray detection; X-ray detectors;