• DocumentCode
    941354
  • Title

    Development and Testing of a 2450-MHz Lens Applicator for Localized Microwave Hyperthermia (Short Paper)

  • Author

    Nikawa, Yoshio ; Kikuchi, Makoto ; Mori, Shinsaku

  • Volume
    33
  • Issue
    11
  • fYear
    1985
  • fDate
    11/1/1985 12:00:00 AM
  • Firstpage
    1212
  • Lastpage
    1216
  • Abstract
    A new type of applicator with a convergent lens for localized microwave hyperthermia is developed. A lens applicator of direct contact type was designed to conduct actual and progressive experiments with phantoms of simulated fat and muscle tissues heated at 2450 MHz. The experimental results showed that the heating power penetration depth increased 40 percent with this applicator as compared to a simple rectangular waveguide applicator with the same size aperture that had generally been used for microwave hyperthermia. Our applicator had a concave-shaped aperture and was designed to contact well with the heating medium whose shape was cylindrical like a human body.
  • Keywords
    Apertures; Applicators; Electromagnetic heating; Hyperthermia; Imaging phantoms; Lenses; Muscles; Optical design; Rectangular waveguides; Testing;
  • fLanguage
    English
  • Journal_Title
    Microwave Theory and Techniques, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9480
  • Type

    jour

  • DOI
    10.1109/TMTT.1985.1133196
  • Filename