    Understanding electrosurgical unit perturbations in order to address hospital operating room electromagnetic compatibility

  • Author

    Fernández-Chimeno, Mireya ; Quílez, Marcos ; Silva, Ferran

  • Author_Institution
    Electron. Eng. Dept., Tech. Univ. of Catalonia, Barcelona, Spain
  • Volume
    53
  • Issue
    6
  • fYear
    2006
  • fDate
    6/1/2006 12:00:00 AM
  • Firstpage
    1206
  • Lastpage
    1209
  • Abstract
    We analyze the radiated emissions from a low power electrosurgical unit(ESU). Measurements at a 3-meter distance are performed in order to find a suitable measurement setup intended to reproduce realistically the real working ESU behavior in a test lab. The broad-band and narrow-band characteristics of the perturbation are studied in order to address the other equipment immunity.
  • Keywords
    bioelectric phenomena; biomagnetism; biomedical equipment; surgery; 3 m; electrosurgical unit perturbations; hospital operating room electromagnetic compatibility; Biomedical measurements; Electromagnetic compatibility; Electromagnetic interference; Electromagnetic measurements; Hospitals; IEC standards; Magnetic field measurement; Performance evaluation; Power generation; Testing; ESU electromagnetic emission measurement setup; Electrosurgical unit EMI measurements; operating room electromagnetic compatibility; Electromagnetic Fields; Electrosurgery; Equipment Failure; Equipment Failure Analysis; Equipment Safety; Operating Rooms; Radiation Dosage; Radiometry;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2006.873389
  • Filename
    1634517