DocumentCode
942421
Title
Use of low-background germanium detectors to preselect high-radiopurity materials intended for constructing advanced ultralow-level detectors
Author
Arthur, R.J. ; Reeves, J.H. ; Miley, H.S.
Author_Institution
Pacific Northwest Lab., Richland, WA, USA
Volume
35
Issue
1
fYear
1988
Firstpage
582
Lastpage
585
Abstract
The use of ultra-low-background germanium-diode gamma-ray spectrometers to measure the concentrations of radionuclides in a variety of materials is described. Bulk shielding materials that were examined included contemporary lead, 150-year-old lead, 400-year-old lead, pre-1945 battleship steel, and electrorefined copper. Typically, the samples were constructed such that a thickness of 5 cm or more of material completely surrounded the detector except for the end connected to the liquid nitrogen dewar. Several experiments were performed to gauge the effects of shielding materials on the observed cosmogenic neutron flux. Another experiment was performed at ground level using one of the germanium diodes fabricated with superclean materials (copper can and cryostat, custom solder, indium o-ring, etc.) and 5-, 10-, and 15-cm-thick low-activity passive Pb shielding to determine the effects of bulk shielding on the observed background. The results are presented and discussed.<>
Keywords
elemental semiconductors; gamma-ray spectrometers; germanium; semiconductor counters; semiconductor diodes; Ge diode; advanced ultralow-level detectors; bulk shielding; cosmogenic neutron flux; gamma-ray spectrometers; ground level; high-radiopurity materials; low background detectors; shielding materials; Building materials; Copper; Detectors; Diodes; Germanium; Lead; Neutrons; Nitrogen; Spectroscopy; Steel;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.12791
Filename
12791
Link To Document