• DocumentCode
    943623
  • Title

    Performance measurements of the SP3000/UW time-of-flight positron emission tomograph

  • Author

    Lewellen, Tk ; Bice, An ; Harrison, RL ; Pencke, MD ; Link, JM

  • Author_Institution
    Dept. of Radiol., Washington Univ., Seattle, WA, USA
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    665
  • Lastpage
    669
  • Abstract
    The basic performance of the University of Washington modified Scanditronix SP3000 time-of-flight positron-emission tomograph (PET) has been characterized. Parameters measured include spatial resolution (in-plane and axial), count rate, sensitivity, scatter fraction, and recovery coefficient. Image examples include the Hoffman phantom and F-18 fluoride studies.<>
  • Keywords
    biomedical equipment; computerised tomography; radioisotope scanning and imaging; Hoffman phantom; SP3000/UW time-of-flight positron emission tomograph; Scanditronix; University of Washington; axial; count rate; in-plane; recovery coefficient; scatter fraction; sensitivity; spatial resolution; Attenuation; Collimators; Crystals; Data processing; Detectors; Photomultipliers; Radioactive decay; Radiology; Sensor arrays; Timing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.12808
  • Filename
    12808