• DocumentCode
    944887
  • Title

    New calibration method for a 7-port reflectometer

  • Author

    Luff, G.E. ; Probert, P.J. ; Carroll, J.E.

  • Author_Institution
    Cambridge University, Engineering Department, Cambridge, UK
  • Volume
    134
  • Issue
    7
  • fYear
    1987
  • fDate
    7/1/1987 12:00:00 AM
  • Firstpage
    595
  • Lastpage
    600
  • Abstract
    The first prototype of a real-time swept-frequency network analyser using two 7-port reflectometers for general use in the microwave laboratory has been reported. It can make simultaneous measurements of all four S parameters of a 2-port device. The instrument relies on a calibration of the multiport reflectometers with an effective method which requires a minimum of standards and is mainly independent of frequency. The success of the prototype instrument is partly due to a novel calibration procedure, which exploits the additional information given by using the 7th port, and uses three frequency-independent standards along with one other approximately known termination.
  • Keywords
    calibration; swept-frequency reflectometry; 2-port device; 7-port reflectometer; S parameters; calibration; frequency-independent standards; microwave laboratory; multiport reflectometers; real-time swept-frequency network analyser;
  • fLanguage
    English
  • Journal_Title
    Physical Science, Measurement and Instrumentation, Management and Education - Reviews, IEE Proceedings A
  • Publisher
    iet
  • ISSN
    0143-702X
  • Type

    jour

  • DOI
    10.1049/ip-a-1.1987.0078
  • Filename
    4647950