Physical performance of the latest generation of commercial positron scanner
Author
Spinks, T.J. ; Jones, T. ; Gilardi, M.C. ; Heather, J.D.
Author_Institution
Hammersmith Hospital, London, UK
Volume
35
Issue
1
fYear
1988
Firstpage
721
Lastpage
725
Abstract
A study of some of the basic physical performance characteristics of an advanced commercial positron-emission tomograph (PET) is reported. It consists of eight rings of BGO (bismuth germanate) detectors (512/ring) with removable interring septa and retractable transmission ring sources for each ring. The (transaxial) detector width is 5.6 mm, and the lower limit of the spatial resolution in the reconstructed image is about 5 mm. Investigations were carried out of count-rate linearity, image count recovery with object size, the effect of attenuation correction on quantitation, and scatter fraction.<>
Keywords
biomedical equipment; computerised tomography; radioisotope scanning and imaging; BiGe/sub 5/O/sub 12/; attenuation correction; commercial positron scanner; count-rate linearity; image count recovery; object size; positron-emission tomograph; quantitation; removable interring septa; retractable transmission ring sources; scatter fraction; Character generation; Crystals; Cyclotrons; Detectors; Face detection; Hospitals; Image reconstruction; Optical scattering; Positrons; Spatial resolution;