    Physical performance of the latest generation of commercial positron scanner

  • Author

    Spinks, T.J. ; Jones, T. ; Gilardi, M.C. ; Heather, J.D.

  • Author_Institution
    Hammersmith Hospital, London, UK
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    721
  • Lastpage
    725
  • Abstract
    A study of some of the basic physical performance characteristics of an advanced commercial positron-emission tomograph (PET) is reported. It consists of eight rings of BGO (bismuth germanate) detectors (512/ring) with removable interring septa and retractable transmission ring sources for each ring. The (transaxial) detector width is 5.6 mm, and the lower limit of the spatial resolution in the reconstructed image is about 5 mm. Investigations were carried out of count-rate linearity, image count recovery with object size, the effect of attenuation correction on quantitation, and scatter fraction.<>
  • Keywords
    biomedical equipment; computerised tomography; radioisotope scanning and imaging; BiGe/sub 5/O/sub 12/; attenuation correction; commercial positron scanner; count-rate linearity; image count recovery; object size; positron-emission tomograph; quantitation; removable interring septa; retractable transmission ring sources; scatter fraction; Character generation; Crystals; Cyclotrons; Detectors; Face detection; Hospitals; Image reconstruction; Optical scattering; Positrons; Spatial resolution;