• DocumentCode
    946135
  • Title

    Urgent problems of metrological assurance of optical pulse oximetry

  • Author

    Lovinsky, Leonid S.

  • Author_Institution
    All-Russian Res. Inst. for Opt. & Phys. Meas., Moscow
  • Volume
    55
  • Issue
    3
  • fYear
    2006
  • fDate
    6/1/2006 12:00:00 AM
  • Firstpage
    869
  • Lastpage
    875
  • Abstract
    It has been shown that the present system of metrological assurance of optical pulse oximetry does not provide uniformity and acceptable level of uncertainties in measurements of oxygen saturation in arterial blood. A new method of evaluation of saturation values and uncertainties has been proposed, permitting to reduce measurement uncertainties to the level that meets the up-to-date metrological requirements. Ways of further reduction of saturation measurement uncertainties by a method of optical pulse oximetry have been shown
  • Keywords
    blood; measurement uncertainty; oximetry; oxygen; patient monitoring; arterial blood; measurement uncertainty; optical pulse oximetry; oxygen saturation; Blood; Electromagnetic wave absorption; Infrared sensors; Infrared spectra; Measurement uncertainty; Optical pulses; Optical saturation; Optical sensors; Pulse measurements; Pulse modulation; Blood; measurement; pulse oximetry; saturation; uncertainties;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2006.873784
  • Filename
    1634880