• DocumentCode
    947179
  • Title

    An alternative discriminator design for the Scanditronix SP-3000 time of flight emission tomograph

  • Author

    Lewellen, Tk ; Pencke, MD ; Bice, An ; Harrison, RL

  • Author_Institution
    Dept. of Radio., Washington Univ., Seattle, WA, USA
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    726
  • Lastpage
    729
  • Abstract
    The detector modules for the original versions of the Scanditronix SP3000 positron-emission tomograph (PET) exhibit a consistent nonuniform sensitivity in count rate for paired crystals coupled to single photomultiplier tubes. The detector modules and discriminators used in the original SP3000 machines are described, and possible solutions are examined. The response is corrected by adopting a new discriminator design for the machine.<>
  • Keywords
    biomedical equipment; computerised tomography; radioisotope scanning and imaging; Scanditronix SP-3000 time of flight emission tomograph; consistent nonuniform sensitivity; count rate; detector modules; discriminator design; paired crystals; single photomultiplier tubes; Circuits; Crystals; Decoding; Delay; Detectors; Energy states; Image reconstruction; Optical coupling; Production systems; Timing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.12820
  • Filename
    12820