Title :
Performance evaluation of MMP-II:A second-generation small animal PET
Author :
Correia, John A. ; Burnham, Charles A. ; Kaufman, David ; Brownell, Anna-Liisa ; Fischman, Alan J.
Author_Institution :
Dept. of Radiol., Massachusetts Gen. Hosp.& Harvard Med. Sch., Boston, MA, USA
Abstract :
We have completed construction of a second-generation, single-plane small animal PET instrument based on LSO detectors. The second-generation design addresses some of the limitations in the first-generation. The purpose of the work reported here was to characterize the physical performance of this instrument. Results of the performance measurements include: Spatial resolution=1.25 mm at field center and 1.5 mm at 2 cm radius; point source sensitivity=56 cps/uCi,; scatter fractions of 0.019 and 0.056 in 3.8 and 6 cm diameter cylinders respectively; linearity of reconstructed signal within 5% up to 100 uCi/cc and acceptable dead-time performance up to 25 k true cps. Examples of phantom and animal images are also presented.
Keywords :
biomedical equipment; biomedical imaging; performance evaluation; phantoms; positron emission tomography; 2 cm; 3.8 cm; 6 cm; LSO detectors; animal images; dead-time performance; phantom; point source sensitivity; scatter fractions; second-generation single-plane small animal PET instrument; spatial resolution; Animals; Detectors; Image reconstruction; Instruments; Linearity; Measurement; Positron emission tomography; Scattering; Signal resolution; Spatial resolution;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2003.823058