• DocumentCode
    948028
  • Title

    Image reconstruction algorithm for a SPECT system with a convergent rotating slat collimator

  • Author

    Zeng, Gengsheng L. ; Gagnon, Daniel

  • Author_Institution
    Med. Imaging Res. Lab., Univ. of Utah, Salt Lake City, UT, USA
  • Volume
    51
  • Issue
    1
  • fYear
    2004
  • Firstpage
    142
  • Lastpage
    148
  • Abstract
    In this paper, we suggest the use of a convergent rotating slat collimator in a SPECT system that contains strip-shaped CdZnTe detectors. This CdZnTe imaging device is able to provide better energy resolution than a conventional NaI(Tl) detector. A convergent imaging geometry has a higher detection sensitivity than a parallel imaging geometry when imaging a small object. A novel design method and reconstruction technique are proposed for use in this system.
  • Keywords
    collimators; geometry; image reconstruction; single photon emission computed tomography; CdZnTe imaging device; SPECT system; convergent imaging geometry; convergent rotating slat collimator; detection sensitivity; energy resolution; filtered backprojection; image reconstruction algorithm; parallel imaging geometry; single photon emission computed tomography; strip-shaped CdZnTe detectors; Attenuation; Cameras; Collimators; Detectors; Energy resolution; High-resolution imaging; Image converters; Image reconstruction; Object detection; Scattering;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2003.823042
  • Filename
    1282076