DocumentCode :
949134
Title :
Repeatability of computer-corrected network-analyser measurements of reflection coefficients
Author :
da Silva, E.F. ; McPhun, M.K.
Author_Institution :
Lanchester Polytechnic, Department of Electrical Engineering, Coventry, UK
Volume :
14
Issue :
25
fYear :
1978
Firstpage :
832
Lastpage :
834
Abstract :
Practical results relating to the accuracy of 100 sets of repeated reflection-coefficient measurements using a computer-corrected network analyser are given. Statistical data based on the above sets of measurements are presented. The results show that measurement accuracies of 1% can be predicted with a confidence level of 99%. A method of dealing with the possible occurrence of `rogue¿ measurements is discussed.
Keywords :
automatic test equipment; measurement errors; microwave measurement; network analysers; computer corrected network analyser; reflection coefficients;