Title :
An improved echo planar reconstruction method
Author :
Kashmar, G.C. ; Nalcioglu, O.
Author_Institution :
California Univ., Irvine, CA, USA
Abstract :
In nuclear magnetic resonance (NMR) imaging, the echo planar method is used for fast scanning. Using conventional reconstruction methods, it requires twice as much data as Cartesian coordinate scanning due to sparse sampling on the edges of the Fourier plane. A noniterative reconstruction method is derived that requires only as much data as the Cartesian case. It operates by exactly interpolating bandlimited echo planar data into Cartesian coordinate data. In order to determine the loss in signal-to-noise ratio (SNR) due to faster operation, simulations are made comparing the new algorithm to a control algorithm that uses the conventional amount of data. It is found that with the new method scanning can be done twice as fast with 46% of the conventional SNR.<>
Keywords :
biomedical NMR; patient diagnosis; Cartesian coordinate scanning; Fourier plane; echo planar reconstruction method; edges; imaging; noniterative reconstruction method; nuclear magnetic resonance; signal-to-noise ratio; sparse sampling; Fourier transforms; Image reconstruction; Interpolation; Iterative methods; Magnetic resonance imaging; Nuclear magnetic resonance; Physics; Reconstruction algorithms; Sampling methods; Signal to noise ratio;
Journal_Title :
Nuclear Science, IEEE Transactions on