DocumentCode
949373
Title
Study of facial skin and aural temperature
Author
Ng, Eddie Y k ; Muljo, Wiryani ; Wong, B. Stephen
Author_Institution
Sch. of Mechanical & Aerosp. Eng., Nanyang Technol. Univ., Singapore
Volume
25
Issue
3
fYear
2006
Firstpage
68
Lastpage
74
Abstract
This article aims to study the correlations between the facial skin surface and contact aural temperatures using two types of IR imagers: 1) IRTRS with only color display mode and 2) ThermalVision900 (TVS900) with direct temperature readings. Results showed that the TVS900 imager Type 2 has good capability in reproducing readings. The noncontact ear temperature was more closely related to contact measured aural (tympanic) reading after running, drinking a hot beverage, and normal conditions. However, the correlation between ear (contact) and eye (noncontact) temperatures taken by IRTRS was relatively lower than what was expected. It is thus essential to give more attention on the device calibration before performing the data collection. The procedures previously suggested allow for more systematic and scientific ways of detecting feverish persons suspected of SARS. The future development of a computerized data processing method is, however, preferable so that less manual work is involved. In the statistical analysis, a setting of 35.5°C was found suitable for an IRTRS that is potentially workable.
Keywords
biomedical optical imaging; biothermics; calibration; diseases; ear; eye; infrared imaging; skin; statistical analysis; IR imagers; IRTRS; SARS; TVS900; ThermalVision900; color display mode; computerized data processing; contact aural temperature; device calibration; direct temperature readings; eye temperatures; facial skin surface temperature; feverish persons; noncontact ear temperature; statistical analysis; tympanic membrane temperature; Birds; Diseases; Displays; Face; Humans; Influenza; Infrared imaging; Optical imaging; Skin; Temperature dependence; Face; Humans; Image Processing, Computer-Assisted; Infrared Rays; Skin; Skin Temperature; Thermometers;
fLanguage
English
Journal_Title
Engineering in Medicine and Biology Magazine, IEEE
Publisher
ieee
ISSN
0739-5175
Type
jour
DOI
10.1109/MEMB.2006.1636353
Filename
1636353
Link To Document