DocumentCode :
955449
Title :
Two-channel high-pressure helium-3 scintillation neutron detector
Author :
Arodzero, A. ; Bolozdynya, A. ; Bolotnikov, A. ; Proctor, A. ; Richards, J.
Author_Institution :
Constellation Technol. Corp., Largo, FL, USA
Volume :
51
Issue :
2
fYear :
2004
fDate :
4/1/2004 12:00:00 AM
Firstpage :
322
Lastpage :
327
Abstract :
We have developed a two-channel, high-pressure 3He gas scintillation detector that operates in coincidence mode. The detector consists of two 16-mm diameter avalanche photodiodes viewing a 7-mm gap between them and mounted inside a stainless steel shell with 1.5-mm thick input windows. The gap is filled with a 3He+0.5%Xe gas mixture pressurized at 35 bar. The detector rejects signals generated by radiation interacting directly with the photodiodes and works as a "wall-less" detector, sensitive to interactions occurring only in the gas. The detector shows axial and radial position sensitivity to thermal neutrons within the gap between photodiodes. The position sensitivity is observed in good agreement with computer simulations. Applications are discussed.
Keywords :
avalanche photodiodes; gas scintillation detectors; neutron detection; nuclear electronics; position sensitive particle detectors; 16 mm; 35 bar; 7 mm; 3He+Xe pressurised gas mixture; avalanche photodiodes; axial position sensitivity; coincidence mode operation; computer simulation; input window; radial position sensitivity; radiation interaction; stainless steel shell; thermal neutrons; two-channel high-pressure 3He gas scintillation neutron detector; wall-less detector; Avalanche photodiodes; Computer simulation; Helium; Neutrons; Radiation detectors; Scintillation counters; Signal detection; Signal generators; Solid scintillation detectors; Steel;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2004.825091
Filename :
1284770
Link To Document :
بازگشت