Abstract
Theoretical considerations for the design of an isotropic combined electric and magnetic radio-frequency field probe are presented. The probe is intended for simultaneous measurements of hazardous electric and magnetic fields under near-field conditions in the frequency range from 10 MHz to 300 MHz. The probe consists of six dipoles and six loops placed on all six sides of a small cube, thus providing a well defined electrical center due to the symmetry. Other significant advantages are improvement of the isotropicity, reduction of the detection errors associated with asymmetrical waveforms, and reduction of the electric field coupling to the magnetic field sensors. Theoretical findings are in agreement with the measured results of a constructed experimental probe
Keywords
electric field measurement; electric sensing devices; magnetic field measurement; probes; 10 to 300 MHz; RF; asymmetrical waveforms; detection errors; dipoles; electric field coupling; electric near field; error compensation; experimental probe; hazardous fields; isotropicity; loops; magnetic field sensors; magnetic near field; radio-frequency field probe; simultaneous measurements; Antenna measurements; Couplings; Electric variables measurement; Geometry; Magnetic field measurement; Magnetic sensors; Magnetic separation; Probes; Radio frequency; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.245653
Filename
245653
Link To Document