Title :
Voltage ratio technique used with the national RF standard attenuation measuring equipment at the PTB
Author :
Rühle, Wolfgang ; Stumpe, Dieter
Author_Institution :
Physikalisch Tech. Bundesanstalt, Braunschweig, Germany
fDate :
12/1/1993 12:00:00 AM
Abstract :
With a new measuring system, the attenuation range of the national RF standard attenuation measuring equipment at the Physikalisch-Technische Bundesanstalt (PTB), Germany, has been extended to 60 db in the waveguide bands R 220 and R 320. Most of the system is automated to shorten the calibration and measurement time. The total uncertainty of measurement is estimated to be 0.0027 dB per 10 dB attenuation
Keywords :
attenuation measurement; calibration; measurement standards; microwave measurement; 18 to 40 GHz; 60 dB; Germany; PTB; Physikalisch-Technische Bundesanstalt; R 220; R 320; attenuation measurement; calibration; measurement time; national RF standard; uncertainty; waveguide bands; Attenuation measurement; Calibration; Couplers; Dynamic range; Frequency measurement; Frequency synthesizers; Measurement standards; Power measurement; Radio frequency; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on