Abstract
Unbalance between electrode-skin impedances is a major problem in biopotential recordings, leading to increased power-line interference. This paper proposes a simple, direct method to measure that unbalance at power-line frequency (50-60 Hz), thus allowing the determination of actual recording conditions for biopotential amplifiers. The method is useful in research, amplifier testing, electrode design and teaching purposes. It has been experimentally validated by using both phantom impedances and real electrode-skin impedances.
Keywords
bioelectric phenomena; biomedical electrodes; biomedical measurement; electric impedance; skin; 50 to 60 Hz; biopotential amplifiers; biopotential recording; electrode-skin impedance unbalance measurement; phantom impedances; power-line interference; Biomedical measurements; Electrodes; Electromagnetic interference; Frequency measurement; Impedance measurement; Instruments; Power amplifiers; Power measurement; Testing; Voltage; Biopotential amplifier; electrode-skin impedance; power-line interference; Amplifiers; Artifacts; Electric Impedance; Electricity; Electrodes; Equipment Design; Equipment Failure Analysis; Reproducibility of Results; Sensitivity and Specificity; Skin Physiology;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2006.875714
Filename
1643419
Link To Document