• DocumentCode
    979317
  • Title

    Power-Efficient Impedance-Modulation Wireless Data Links for Biomedical Implants

  • Author

    Mandal, S. ; Sarpeshkar, R.

  • Author_Institution
    Analog VLSI & Biol. Syst. Group, Massachusetts Inst. of Technol. (MIT), Cambridge, MA
  • Volume
    2
  • Issue
    4
  • fYear
    2008
  • Firstpage
    301
  • Lastpage
    315
  • Abstract
    We analyze the performance of wireless data telemetry links for implanted biomedical systems. An experimental realization of a bidirectional half-duplex link that uses near-field inductive coupling between the implanted system and an external transceiver is described. Our system minimizes power consumption in the implanted system by using impedance modulation to transmit high-bandwidth information in the uplink direction, i.e., from the implanted to the external system. We measured a data rate of 2.8 Mbps at a bit error rate (BER) of <10-6 (we could not measure error rates below 10-6 ) and a data rate of 4.0 Mbps at a BER of 10-3. Experimental results also demonstrate data transfer rates up to 300 kbps in the opposite, i.e., downlink direction. We also perform a theoretical analysis of the bit error rate performance. An important effect regarding the asymmetry of rising and falling edges that is inherent to impedance modulation is predicted by theory and confirmed by experiment. The link dissipates 2.5 mW in the external system and only 100 muW in the implanted system, making it among the most power-efficient inductive data links reported. Our link is compatible with FCC regulations on radiated emissions.
  • Keywords
    biomedical telemetry; error statistics; prosthetics; bidirectional half-duplex link; biomedical implant; bit error rate; data transfer rates; inductive coupling; power-efficient impedance-modulation; uplink direction; wireless data links; wireless data telemetry links; Biomedical measurements; Bit error rate; Downlink; Energy consumption; Error analysis; Impedance; Implants; Performance analysis; Telemetry; Transceivers; Impedance modulation; implants; load modulation; low power; near-field; telemetry; wireless;
  • fLanguage
    English
  • Journal_Title
    Biomedical Circuits and Systems, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1932-4545
  • Type

    jour

  • DOI
    10.1109/TBCAS.2008.2005295
  • Filename
    4667645