DocumentCode
986602
Title
Performance Improvement of a DSP-Based Digital Integrator for Magnetic Measurements at CERN
Author
Arpaia, Pasquale ; Inglese, Vitaliano ; Spiezia, Giovanni
Author_Institution
Dept. of Eng., Univ. of Sannio, Benevento
Volume
58
Issue
7
fYear
2009
fDate
7/1/2009 12:00:00 AM
Firstpage
2132
Lastpage
2138
Abstract
An improved self-calibrating digital instrument for online flux measurements on superconducting magnets for particle accelerators prototyped at the European Organization for Nuclear Research (CERN), in cooperation with the University of Sannio, is presented. The new instrument reduces the online flux analysis time down to 4.00 mus within a resolution of 50 ns. Furthermore, the typical signal-to-noise-and-distortion ratio (SINAD) is significantly improved with respect to the previous version and to the instruments that are currently in use. Details about the design, online measurement principle, and results of the metrological characterization of the new instrument are provided.
Keywords
digital circuits; digital integrated circuits; high energy physics instrumentation computing; integrating circuits; nuclear electronics; storage rings; superconducting magnets; CERN; DSP-based digital integrator; European Organization for Nuclear Research; magnetic measurements; metrological characterization; online flux measurements; particle accelerators; self-calibrating digital instrument; signal-to-noise-and-distortion ratio; superconducting magnets; Accelerator measurement systems; digital measurements; integrators; magnetic variables measurement; signal processing;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2008.2006723
Filename
4671089
Link To Document