DocumentCode :
989520
Title :
CMOS retinal prosthesis with on-chip electrode impedance measurement
Author :
Uehara, A. ; Ng, D.C. ; Kagawa, K. ; Tokuda, T. ; Ohta, J. ; Nunoshita, M.
Author_Institution :
Nara Inst. of Sci. & Technol., Ikoma Nara, Japan
Volume :
40
Issue :
10
fYear :
2004
fDate :
5/13/2004 12:00:00 AM
Firstpage :
582
Lastpage :
584
Abstract :
A retinal prosthesis device with built-in self-test capability is proposed and demonstrated. The measurement of electrode impedance as a self-test is achieved without increasing the chip area by employing analogue multiplexers to allow the electrodes to be used for both stimulation of retinal cells and measurement of impedance. Measurement is performed using a four-terminal method to ensure good accuracy. A prototype stimulus chip with 16×16 channels is fabricated using standard 0.6 μm CMOS technology, and is demonstrated to provide self-test functionality with error of as little as 0.05% in the frequency range of 100 Hz to 100 kHz.
Keywords :
CMOS integrated circuits; biomedical electrodes; biomedical electronics; built-in self test; electric impedance; eye; prosthetics; 0.6 micron; 100 Hz to 100 kHz; CMOS retinal prosthesis; CMOS technology; electrodes; four terminal method; on-chip electrode impedance measurement; retinal cells;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el:20040410
Filename :
1300217
Link To Document :
بازگشت