DocumentCode :
990042
Title :
Radiometer input circuit requirements for microwave thermography
Author :
Land, D.V.
Author_Institution :
University of Glasgow, Department of Natural Philosophy, Glasgow, UK
Volume :
19
Issue :
24
fYear :
1983
Firstpage :
1040
Lastpage :
1042
Abstract :
Radiometer system performance requirements and problems relevant to the technique of medical microwave thermography are briefly reviewed. The Dicke-type radiometer input circuit normally used for this application is analysed to determine the factors limiting measurement performance. It is shown that input circuit losses combined with aerial to body impedance mismatch can cause significant measurement errors.
Keywords :
biomedical measurement; microwave measurement; radiometers; Dicke-type radiometer; aerial to body impedance mismatch; input circuit losses; medical microwave thermography;