DocumentCode :
990381
Title :
A Novel Method to Read Remotely Resonant Passive Sensors in Biotelemetric Systems
Author :
Pichorim, Sérgio Francisco ; Abatti, Paulo José
Author_Institution :
Fed. Univ. of Technol., Parana
Volume :
8
Issue :
1
fYear :
2008
Firstpage :
6
Lastpage :
11
Abstract :
Resonant passive sensors composed by resistive, inductive, and capacitive (RLC) simple sensors are widely used in biotelemetric systems. In this paper, a novel method to read remotely these RLC sensors is presented. The developed method is based on the simultaneous application of three excitation signals of same amplitudes, set at different frequencies, to determine remotely the RLC sensor resonance frequency and quality factor . Theoretical analysis and experimental results are also presented.
Keywords :
Q-factor; biomedical telemetry; biosensors; biotelemetric system; quality factor; remotely resonant passive sensor; sensor resonance frequency; Biosensors; Capacitive sensors; Coils; In vivo; Q factor; Radio frequency; Resonance; Resonant frequency; Sensor systems; Telemetry; Biomedical instrumentation; biomedical telemetry; implantable sensors; quality factor; resistive, inductive, and capacitive (RLC) passive sensor; resonance frequency; wireless telemetry;
fLanguage :
English
Journal_Title :
Sensors Journal, IEEE
Publisher :
ieee
ISSN :
1530-437X
Type :
jour
DOI :
10.1109/JSEN.2007.912386