DocumentCode :
991924
Title :
Optimized configurations for passively shielded magnetic resonance imaging magnets
Author :
Kalafala, A.K.
Author_Institution :
General Electric Corp. Res. & Dev., Schenectady, NY, USA
Volume :
29
Issue :
2
fYear :
1993
fDate :
3/1/1993 12:00:00 AM
Firstpage :
1240
Lastpage :
1244
Abstract :
A design approach is presented for the development of passive shields for coil configurations, with particular emphasis on those used for magnetic resonance imaging (MRI). The approach is based on minimizing the weight of the shield by ensuring that the shield material is everywhere near saturation. A mathematical model is developed to describe the shielding efficiency of a cylindrical shield as a function of shield geometry and type of shield material. The model is used to develop a possible multiple-shield configuration for a 1.5-T MRI magnet
Keywords :
biomedical NMR; coils; electromagnets; magnetic shielding; nuclear magnetic resonance imaging; 1.5 T; MRI; coil configurations; cylindrical shield; design; magnetic resonance imaging; mathematical model; medical imaging; multiple-shield configuration; optimised configuration; passively shielded magnetic resonance imaging magnets; saturation; shield geometry; shield material; shielding efficiency; weight minimisation; Coils; Geometry; Helium; Impedance; Magnetic field measurement; Magnetic materials; Magnetic resonance imaging; Magnetic shielding; Magnets; Mathematical model;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/20.250628
Filename :
250628
Link To Document :
بازگشت