    Charge division read-out in a 1.8-m-long wire chamber

  • Author

    Lan, K.A. ; Hungerford, E.V. ; Johnson, K. ; Ambrose, D. ; Coffey, P. ; Wang, J.J.

  • Author_Institution
    Dept. of Phys., Houston Univ., TX, USA
  • Volume
    42
  • Issue
    4
  • fYear
    1995
  • fDate
    8/1/1995 12:00:00 AM
  • Firstpage
    1430
  • Lastpage
    1438
  • Abstract
    This paper reports on the design and the test results of a charge division electronic read-out system for a charged particle proportional wire detector. Electronic calibration and probable error analysis of the system are also discussed. Position resolution for the system has been measured to be 4.8-mm FWHM on a wire of 1.8-m length (0.26%, relative)
  • Keywords
    calibration; detector circuits; nuclear electronics; proportional counters; 1.8 m; charge division read-out; charged particle proportional wire detector; electronic calibration; position resolution; probable error analysis; Cathodes; Charge measurement; Current measurement; Detectors; Energy resolution; Mesons; Position measurement; Pulse shaping methods; Signal processing; Wire;