Title :
Investigation of the count rate performance of General Electric Advance positron emission tomograph
Author :
Lewellen, T.K. ; Kohlmyer, S.G. ; Miyaoka, R.S. ; Schubert, S. ; Stearns, C.W.
fDate :
8/1/1995 12:00:00 AM
Abstract :
A series of 2D and 3D count rate measurements have been carried out on the Advance PET system. The measurements include: NEMA standard protocols; simultaneous imaging of two small vials containing [0-15]water and [F-18]; a cardiac insert placed in a water filled elliptical tank; two different head sized cylindrical tanks; and a Hoffman 3D phantom. The data were processed with and without deadtime and scatter corrections. Plots were made of the total count rate, randoms, trues+scatter, scatter corrected trues, and noise equivalent count rate as a function of total activity in the machine and as the μCi/cc in the phantoms. The accuracy of deadtime corrections was also measured. In 2D mode, the NEMA standard protocol resulted in maximum `trues´ count rates of 540,000 cps in high sensitivity mode. In 3D mode the head sized cylindrical phantoms yielded a maximum trues rate of approximately 950,000 cps. Recovery of the water half-life was accurate with up to 50 mCi for the cardiac insert in 2D mode
Keywords :
cardiology; positron emission tomography; prosthetics; 0.05 Ci; F; General Electric Advance PET; H2O; Hoffman 3D phantom; NEMA standard protocols; cardiac insert; count rate measurements; count rate performance; deadtime corrections; head sized cylindrical phantoms; head sized cylindrical tanks; high sensitivity mode; maximum trues rate; noise equivalent count rate; positron emission tomograph; scatter corrections; simultaneous imaging; small vials; water filled elliptical tank; water half-life recovery; Circuits; Detectors; Event detection; Imaging phantoms; Magnetic heads; Positron emission tomography; Protocols; Radioactive decay; Scattering; System testing;
Journal_Title :
Nuclear Science, IEEE Transactions on