Title :
Image denoising based on multiscale singularity detection for cone beam CT breast imaging
Author :
Zhong, Junmei ; Ning, Ruola ; Conover, David
Author_Institution :
Dept. of Radiol., Rochester Univ., NY, USA
fDate :
6/1/2004 12:00:00 AM
Abstract :
It was recently reported that the real-time flat panel detector-based cone-beam computed tomography (CBCT) breast imaging can help improve the detectability of small breast tumors with an X-ray dose comparable to that of the conventional mammography. In this paper, an efficient denoising algorithm is proposed to further reduce the X-ray exposure level required by a CBCT scan to acquire acceptable image quality. The proposed wavelet-based denoising algorithm possesses three significant characteristics: 1) wavelet coefficients at each scale are classified into two categories: irregular coefficients, and edge-related and regular coefficients; 2) noise in irregular coefficients is reduced as much as possible without producing artifacts to the denoised images; and 3) for the edge-related and regular coefficients, if they are at the first decomposition level, they are further denoised, otherwise, no modifications are made to them so as to obtain good visual quality for diagnosis. By applying the proposed denoising algorithm to the filtered projection images, the X-ray exposure level necessary for the CBCT scan can he reduced by up to 60% while obtaining clinically acceptable image quality. This denoising result indicates that in the clinical application of CBCT breast imaging, the patient radiation dose can be significantly reduced.
Keywords :
computerised tomography; diagnostic radiography; image classification; image denoising; image representation; least mean squares methods; mammography; medical image processing; tumours; wavelet transforms; X-ray exposure; acceptable image quality; breast imaging; breast phantom; cone-beam computed tomography; edge-related coefficients; filtered projection images; image denoising; irregular coefficients; mammography; multiscale image representation; multiscale singularity detection; real-time flat panel detector-based tomography; regular coefficients; small breast tumor detectability; wavelet-based denoising algorithm; Breast tumors; Computed tomography; Image denoising; Image quality; Mammography; Noise reduction; Optical imaging; X-ray detection; X-ray detectors; X-ray imaging; Algorithms; Breast Diseases; Breast Neoplasms; Calcinosis; Humans; Mammography; Phantoms, Imaging; Radiographic Image Enhancement; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Stochastic Processes; Tomography, X-Ray Computed;
Journal_Title :
Medical Imaging, IEEE Transactions on
DOI :
10.1109/TMI.2004.826944