Title :
Automated measurement of noisy voltages with a preset confidence level
Author_Institution :
CSIRO, National Measurement Laboratory, Sydney, Australia
Abstract :
A logical way of measuring voltages having varying amounts of noise present, under computer control, has been achieved by evaluating the running standard deviation of the mean in real time and stopping the measurements when a preset threshold has been reached.
Keywords :
computerised instrumentation; noise; voltage measurement; automated measurement; computer control; noisy voltages; preset confidence level;
Journal_Title :
Electronics Letters
DOI :
10.1049/el:19770170