DocumentCode :
998005
Title :
Optimization of the internal admittance load of an inductive voltage divider for low ratio error
Author :
Lu, Zuliang ; Klonz, Manfred ; Bergeest, Rainer
Author_Institution :
Nat. Inst. of Metrol., Beijing, China
Volume :
39
Issue :
3
fYear :
1990
fDate :
6/1/1990 12:00:00 AM
Firstpage :
463
Lastpage :
466
Abstract :
The distributed capacitances between the strands of an inductive voltage divider degrade the voltage ratio. A calculation scheme to convert the distributed capacitances into equivalent concentrated capacitances at the taps is given. An optimized connection scheme for the strands is calculated to produce concentrated capacitances which are as equal and as small as possible. Capacitors are added at the taps for optimum equalization. The relative error of the output voltage for a 120-V:10-V divider has been reduced to less than 1×10-7 in phase and less than 1 μrad in quadrature, at a frequency of 50 Hz
Keywords :
electric admittance; measurement standards; voltage dividers; 120 to 10 V; 50 Hz; distributed capacitances; equivalent concentrated capacitances; inductive voltage divider; internal admittance load; relative error; voltage ratio; Admittance; Capacitance measurement; Capacitors; Copper; Degradation; Frequency conversion; Helium; Low voltage; Magnetic materials; Metrology;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.106273
Filename :
106273
Link To Document :
بازگشت