DocumentCode :
999888
Title :
Characterisation of high frequency focused ultrasonic transducers using modulation transfer function: concept and experimental approach
Author :
Shiloh, K. ; Som, A.K. ; Bond, L.J.
Author_Institution :
Soreq NCR, Yaune, Israel
Volume :
138
Issue :
4
fYear :
1991
fDate :
7/1/1991 12:00:00 AM
Firstpage :
205
Lastpage :
212
Abstract :
The concept of the modulation transfer function, which is used to evaluate high quality optical systems, components and materials, has been adapted for the first time in a systematic way in ultrasonics to characterise high frequency focused transducers. The additional complexity of ultrasonics compared with optics has been analysed and treated. A 2-D MTF characteriser is presented, as a counterpart to the 1-D function in optics. An example of a complete experimental procedure providing the full performance of focused transducer is presented.
Keywords :
optical transfer function; ultrasonic transducers; 2-D MTF; MTF; acoustic microscopy; high frequency focused ultrasonic transducers; modulation transfer function;
fLanguage :
English
Journal_Title :
Science, Measurement and Technology, IEE Proceedings A
Publisher :
iet
ISSN :
0960-7641
Type :
jour