عنوان مقاله :
ساخت اتاقك يونش چاهك دار مرتبط با هواي آزاد و ارزيابي محاسباتي و اندازه گيري پارامترهاي عملكردي آن
عنوان به زبان ديگر :
Fabrication of Free Air Well Type Ionization Chamber and Calculational Assessment and Measurement of Its Operational Characteristics
پديد آورندگان :
اربابي، كورش نويسنده پژوهشكده تحقيقات كشاورزي پزشكي و صنعتي-پژوشهگاه علوم و فنون هسته اي-سازمان انرژي اتمي ايران Arbabi, Koroush , رييس علي، غلام رضا نويسنده پژوهشكده كاربرد پرتوها-پژوهشگاه علوم و فنون هسته اي-سازمان انرژي اتمي ايران Raisali, Gholamreza , دهقاني مقدم، معصومه نويسنده معاونت تحقيقات و فناوري هسته اي-سازمان انرژي اتمي ايران Dehghani Moghadam, Masoomeh , غفوري، مصطفي نويسنده پژوهشكده تحقيقات كشاورزي پزشكي و صنعتي-پژوهشگاه علوم و فنون هسته اي-سازمان انرژي اتمي ايران Ghafoori, Mostafa , شاهور، ارژنگ نويسنده پژوهشكده تحقيقات كشاورزي پزشكي و صنعتي-پژ<هشگاه علوم و فنون هسته اي-سازمان انرژي اتمي ايران Shahvar, Arjang , حميدي، سعيد 1339 نويسنده گروه فيزيك-دانشكده علوم پايه-دانشگاه اراك Hamidi, Said
اطلاعات موجودي :
فصلنامه سال 1386 شماره 16
رتبه نشريه :
علمي پژوهشي
تعداد صفحه :
10
از صفحه :
1
تا صفحه :
10
كليدواژه :
چشمه هاي گاماي براكي تراپي , قدرت گرماي هوا , شبيه سازي مونت كارلو , اتاقك يونش چاهكدار , هواي آزاد
چكيده لاتين :
Introduction: Well type ionization chamber is a measuring device which is used to determine the activity of brachytherapy sources. The chamber has a cylindrical volume in which a cylindrical tube is mounted in the middle of the chamber. For the measurements, the brachytherapy sources are transferred to the middle of the tube. Materials and Methods: For designing the well type chamber, the measurement principals of well type chambers were considered and MCNP-4C code as a calculation tool was used. The designed chamber was simulated and the response of the chamber was evaluated. In this investigation, the chamber operational parameters such as operating voltage, leakage current, reproducibility, reference measuring point, recombination and polarization factors as well as response stability for 137Cs, 57Co and 241Am sources were studied. Results: The chamber leakage currents at the operating voltage in comparison to the chamber response for the measurement of the above mentioned sources were negligible. The responses of the fabricated chamber for these sources are reproducible and its reference measurement position for these sources was obtained at 6 cm from the bottom of the chamber. The recombination factor for the well type chamber was negligible and the polarization factor is close to 1. Therefore, these two factors were not considered in the measurements. The reproducibility of the measurements in different intervals shows the stability of the chamber response for each source. Also the results of the chamber current measurement in term of source strength were compared to the response of the simulated chamber for different source positions and energy ranges of the used sources. Discussion and Conclusion: The results show that the measurement of the reference positions for each source in the simulated and fabricated chamber is quite in a good agreement. Regarding the reliable operational properties of the fabricated chamber, this chamber can be used to determine the activities of brachytherapy sources in which the energy ranges are greater than 50 KeV.
سال انتشار :
1386
اطلاعات موجودي :
فصلنامه با شماره پیاپی 17 سال 1386
كلمات كليدي :
#تست#آزمون###امتحان
لينک به اين مدرک :
بازگشت