Title of article :
Simultaneous determination of follicle stimulating hormone and luteinising hormone using a multianalyte immunosensor
Author/Authors :
D.J. Pritchard، نويسنده , , H. Morgan، نويسنده , , JM Cooper، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
6
From page :
251
To page :
256
Abstract :
A technique is described for constructing a multianalyte immunosensor for the quantitative determination of the human gonadotrophin hormones follicle stimulating hormone (FSH) and luteinising hormone (LH). The assay is a based upon the electrochemical detection of two horseradish peroxidase labelled antibodies using a ferrocene mediated system. The detection limits (3σ) were 2.1 U l−1 FSH and 1.8 U l−1 LH. There was negligible cross reactivity. Results obtained with the biosensor showed a good correlation with those obtained using an established clinical diagnostic technique based upon dissociation-enhanced lanthanide fluorometric immunoassay. This is the first report of an electrochemical multianalyte immunosensor for high molecular weight analytes.
Keywords :
Biosensors , Multianalyte , Protein patterning , Antibodies , Hormones
Journal title :
Analytica Chimica Acta
Serial Year :
1995
Journal title :
Analytica Chimica Acta
Record number :
1022717
Link To Document :
بازگشت