Title of article :
A coulometer for determination of residual sulfite following the dechlorination of water
Author/Authors :
Nigama Ekkad، نويسنده , , Calvin O. Huber، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
6
From page :
155
To page :
160
Abstract :
The kinetics of dechlorination dictate that the measurement must be completed in neutral solutions. A portable, analog circuitry-based instrument with direct readout using constant current coulometry with amperometric end point detection was developed and evaluated. An optocoupler and separate power supplies were used for circuit isolation of the detector from the generator. A range of concentrations of 0.015–25 mg SO2−3/1 were determined with a relative standard deviation of 1–2%. Applications involved both iodine and bromine as intermediates. The instrument has also been applied to determinations of arsenic, mercaptans, thiosulfate, copper, cyanide and ammonia for a similar concentration range.
Keywords :
Coulometry , Sulfite , Dechlorination , Mercaptans , Cyanide , Ammonia
Journal title :
Analytica Chimica Acta
Serial Year :
1996
Journal title :
Analytica Chimica Acta
Record number :
1024247
Link To Document :
بازگشت