Title of article
Voltammetric and amperometric study of electrochemical activity of boron-doped polycrystalline diamond thin film electrodes
Author/Authors
Jiali Wu، نويسنده , , Jianzhong Zhu، نويسنده , , Lei Shan، نويسنده , , Nianyi Cheng، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
6
From page
125
To page
130
Abstract
The electrochemical activity of boron-doped polycrystalline diamond thin film (BDF) electrodes was studied by using voltammetry to determine hexacyanoferrate(III/II) in KCl and HClKCl (pH = 1.3) solutions. The preliminary results were compared with those given by glassy carbon (GC) electrodes and proved that the BDF electrodes offered high sensitivity, good precision and extreme stability over a month period. The effects of pH and their ohmic resistance on response current of BDF electrodes were also studied.
Keywords
Voltammetry , Response current , Ohmic resistance , BDF electrode
Journal title
Analytica Chimica Acta
Serial Year
1996
Journal title
Analytica Chimica Acta
Record number
1024280
Link To Document