• Title of article

    Biosensor for phenol based on the direct electron transfer blocking of peroxidase immobilising on silica–titanium Original Research Article

  • Author/Authors

    Simone S. Rosatto، نويسنده , , Lauro T. Kubota، نويسنده , , Graciliano de Oliveira Neto، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    8
  • From page
    65
  • To page
    72
  • Abstract
    Horseradish peroxidase (HRP) was immobilised on silica gel modified with titanium oxide. This material was employed to prepare modified carbon paste electrode. The direct electron transfer of the hydrogen peroxide reduction by HRP was blocked when immobilised on silica–titanium. This biosensor presented a very sensitive response for phenol (1 μmol l−1) at an applied potential of 0 mV vs SCE. The best condition was achieved in phosphate buffer pH 6.8, ratio of hydrogen peroxide/phenol higher than 0.35. The biosensor showed a linear response range between 10 and 50 μmol l−1 of phenol, adjusted by the equation j=−32.8+16.3 [phenol], for n=5 with a correlation coefficient of 0.9995. The response time of the biosensor was about 3 s.
  • Keywords
    Peroxidase , Modified silica , Phenol biosensor
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1999
  • Journal title
    Analytica Chimica Acta
  • Record number