Issue Information :
ماهنامه با شماره پیاپی سال 2005
Pages :
-637
From page :
638
To page :
0
Abstract :
This study aimed to develop a simple and efficient optimized high-performance liquid chromatograph (HPLC) method for simultaneous determination of cyclosporine A (CyA) and its major, partly active metabolites AM1, AM9, AM4N, and AM19 in whole blood from transplant patients using cyclosporine D (CyD) as internal standard. The method used a CN analytical column maintained at 60 (degree)C with hexan-isopropanol (93:7, v/v) as mobile phase; detection was at 212 nm. Linearity for all five compounds was tested in the range of 311500 ng ml^-1 for CyA and of 31-1000 ng ml^-1 for metabolites. The limit of detection was found to be 15 ng ml^-1 for all compounds. This modified, inexpensive method is also suitable for measuring cyclosporine A and metabolite concentrations in routine monitoring of patients undergoing treatment with CyA.