Title of article
Using Mass Spectrometry to Characterize Permeation Half-Life of ICF Targets
Author/Authors
Nikroo، A. نويسنده , , Jaquez، J. S. نويسنده , , Alfonso، E. L. نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
-772
From page
773
To page
0
Abstract
A mass spectrometer-based system was developed to measure the collected gas permeating through the shell wall of a spherical ICF target. The resultant ion current was used to calculate the materialʹs permeation half-life. This novel technique is simple, non-destructive, and suitable for measuring targets with short or long half-lives, i.e., very permeable or impermeable target materials, typically in less than 30 min. The technique is suitable for measuring permeation at ambient and elevated temperatures. The results acquired with the mass spectrometer measurements agreed well with those from destructive techniques.
Keywords
fast ignition , fast heating , cone shell
Journal title
FUSION SCIENCE AND TECHNOLOGY
Serial Year
2006
Journal title
FUSION SCIENCE AND TECHNOLOGY
Record number
112778
Link To Document